Size |
55 x 160 x 25 mm (2.2 x 6.3 x 1.0 in) |
Weight |
140 g (battery pack included, 4.9 oz) |
Turbines |
|
Battery pack |
Rechargeable lithium-ion 3.7 V, 1100 mAh |
Current Consumption |
~20-30 mA (during testing) |
Charge Batteries |
Voltage=5 V DC
Current = minimum 500 mA
Input current = 100VAC - 240 VAC
Connector: micro USB type B compliant with EN 60601-1 |
Autonomy |
50 hours |
Connectivity |
USB 2.0 |
Display |
monochrome LCD, 160 x 80 pixels |
Keyboard |
6-key membrane |
Mouthpiece |
Ø 30 mm (1.18 in) |
Type of electrical protection |
Powered internally |
Safety level due to shock hazard |
Type BF device |
IP protection level |
IPX1 |
Terms of use |
Device for continuous use |
Storage conditions |
Temp: MIN -20°C, MAX+60°C Humidity: MIN 10% RH; MAX 95%RH |
Operating conditions |
Temp: MIN +10°C, MAX +40°C Humidity: MIN 10% RH, MAX 95%RH |
Sensor |
two-way digital turbine |
Flow range |
±16L/s |
Volume accuracy |
±2.5% or 50mL |
Flow accuracy |
±5% or 200 mL/s |
Dynamic resistance |
<0.5 cm H2O/L/s |
Temperature sensor |
semiconductor (0-45°C) |
Available tests |
FVC, VC, IVC, POST |
Measured parameters (Stand-alone mode) |
*FVC, *FEV1, *PEF, FVC, FEV1, FEV1/FVC, PEF, FEF25-75, FET, BEV,VC, IVC, IC, ERV, ELA
*Best values
|
Measured parameters (via MIR Spiro software) |
FVC, FEV1, FEV1%, PEF, FEF25–75, FET, VC, IVC, IC, Extr. Vol, ELA ERV,MVV
|
Memory capacity |
more than 10,000 tests |