Re-Usable Mouthpiece Sterilization Instructions
|
A-M Systems Mouthpieces and Respiratory Connectors Recommended Cleaning and Sterilization Instructions
[Download PDF Version]
|
WARNING: A-M Systems thermoplastic, reusable, products are supplied non-sterile and should be sterilized before first use and after each patient use.
Recommended Cleaning Instructions:
- Clean the products using one of the following validated processes:
- Method 1:
- Thoroughly rinse products under tap water for a minimum of one minute to remove gross soil and debris.
- Prepare cleaning solution using Enzol®*, or equivalent, and lukewarm tap water. Prepare the solution according to the manufacturer’s recommendation.
- Using a soft bristled brush, dip the brush into the prepared detergent and scrub products for a minimum of one minute each, paying particular attention to crevices and other hard to reach areas.
- Once the products have been scrubbed, thoroughly rinse in reverse osmosis / deionized (RO/DI) water for a minimum of one minute to remove excess detergent.
- Immediately after rinsing, drain off excess water from products.
- Blow-dry the products using compressed air, especially in crevices and other hard to reach areas.
- Wipe and thoroughly dry products using a clean, lint-free, cloth.
- Visually inspect products to ensure all visible soil has been removed.
- Method 2:
- Thoroughly rinse products under tap water for a minimum of one minute to remove gross soil and debris.
- Using a sonication unit, prepare a solution consisting of Renu-Klenz™**, or equivalent, and lukewarm tap water. Prepare the solution according to the manufacturer’s recommendation.
- Place products, fully immersed, in the detergent solution and allow to sonicate for ten minutes.
- Remove products from sonication unit and thoroughly rinse in reverse osmosis / deionized (RO/DI) water for a minimum of one minute to remove excess detergent.
- Immediately after rinsing, drain off excess water from products.
- Blow-dry the products using compressed air, especially in crevices and other hard to reach areas.
- Wipe and thoroughly dry products using a clean, lint-free, cloth.
- Visually inspect products to ensure all visible soil has been removed.
Recommended SterilizationInstructions:
- Manually clean the products before sterilization according to attached cleaning instructions.
- Sterilize the products using the following validated process:
- Individually pouch the products
- Use a Prevacuum (Steam under Pressure) Cycle of 4 Minutes at 132 °C with a minimum 20-minute Drying Time and 3 Preconditioning Pulses.
- After completion of the Prevacuum Cycle, immediately remove the products from the autoclave and store them in a dry/contamination free environment prior to use.
References:
- "Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: FDA Reviewer Guidance", Rockville, MD, FDA Center for Devices and Radiological Health, Office of Device Evaluation, April 1996
- Association for the Advancement of Medical Instrumentation. "Designing, Testing, and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers". AAMI TIR No. 12-200
- Arlington (Vir.): AAMI Technical Information Report.
- Association for the Advancement of Medical Instrumentation. "A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices". AAMI TIR No. 30-200
- Arlington (Vir.): AAMI Technical Information Report.
- Association for the Advancement of Medical Instrumentation. "Safe Handling and Biological Decontamination of Medical Devices in Health Care Facilities and in Nonclinical Settings", ANSI/AAMI ST35-2003, Arlington (Vir.) AAMI, 2003 American National Standard.
* Enzol® is a trademark of Johnson & Johnson Medical Products, Inc.
** Renu-Klenz™ is a trademark of STERIS Corporation.